Strain Gauge Load Cell: How It Works, Wiring, mV/V and Types

By editor 8 min read
Wheatstone bridge wiring of a strain gauge load cell

A strain gauge load cell turns force into a small voltage. Inside, four strain gauges are bonded to a metal spring element and wired as a Wheatstone bridge. Load bends the element, the gauges change resistance, and the bridge output rises in proportion to the force. Output is rated in mV per volt of excitation, typically 1 to 3 mV/V. A 2 mV/V cell on 10 V gives 20 mV at full load, so 200 kg on a 500 kg cell gives 8 mV. Strain gauge cells are the most common industrial type because they are accurate, versatile and cost-effective (Wikipedia).

1–3 mV/V
typical rated output
10 V
usual excitation
350 Ω
usual gauge resistance
≈ 2
foil gauge factor

Load Cell vs Strain Gauge

People often use the two terms for the same thing, but they are different parts. A strain gauge is very fine wire or foil laid out in a grid on a flexible backing. Stretching it makes it longer and thinner, so its resistance rises. Compressing it does the opposite. A load cell is the finished force transducer. It has an aluminum, alloy steel or stainless steel body (the spring element), several gauges bonded where strain is highest, a bridge circuit, sealing and a cable. It is calibrated so that output tracks load.

Strain gaugeLoad cell
What it isSensing element (foil or wire grid)Complete force transducer
MeasuresStrain (relative length change)Force or weight
OutputTiny change in resistancemV/V bridge signal (or amplified/digital)
Calibrated inGauge factorRated capacity and mV/V
Used forStress analysis, building sensorsScales, tank weighing, press and tension monitoring
IndustrialBasics comparison based on Wikipedia (Strain gauge, Load cell).

How a Strain Gauge Load Cell Works

The change in one gauge is tiny. The gauge factor relates it to strain: GF = (ΔR/R) ÷ ε, and Wikipedia gives a little over 2 for metallic foil gauges. A Wheatstone bridge converts the change to a voltage. Load cells put four gauges in the bridge, alternating between tension and compression, so all four changes add to the signal. With no load the bridge balances and S+ minus S− is about zero. Under load the output grows in proportion to the force.

Vo = (R3/(R3+R4) − R2/(R1+R2)) · VEX     Quarter bridge: Vo ≈ VEX · GF · ε / 4

For a single active gauge with three fixed resistors (a quarter bridge), the output is about VEX × GF × ε / 4. At 10 V, GF 2 and 1,000 microstrain that is only 5 mV. That is why a full bridge with four active gauges is used in load cells, and why the signal needs a high-gain, low-noise amplifier.

Wheatstone bridge wiring of a strain gauge load cellDiagram of four strain gauges in a Wheatstone bridge: tension gauges SG1 and SG3 and compression gauges SG2 and SG4 in alternating arms; excitation E+ red and E− black at top and bottom, signal S+ green and S− white at the sides. Panel lists usual excitation 10 V, rated output 1 to 3 mV/V, gauge resistance 350 ohms and four or six wire cables.Inside a strain gauge load cell: a full Wheatstone bridgeSG1 tensionSG2 compressionSG3 tensionSG4 compressionE+ (red)E− (black)S+ (green)S− (white)Excitationusually 10 V (also 5 V, 20 V)Rated output1 to 3 mV/V at full loadExample2.96 mV/V × 10 V = 29.6 mVGauge resistanceusually 350 Ω (120, 1000 Ω)Wires4-wire, or 6-wire with senseNo loadS+ − S− ≈ 0 VGauges in tension and compression sit in opposite arms, so their changes add at the output.IndustrialBasicsindustrialbasics.com
Figure 1: Bridge arrangement, typical values and the most common wire colours from Wikipedia (Load cell); always confirm colours on the cell’s data sheet.

Excitation, Rated Output and Wiring

  • Excitation: a stabilized voltage across E+ and E−, usually 10 V, sometimes 20 V or 5 V, or lower on battery instruments. Typical maximum is around 15 V.
  • Rated output: mV/V at full rated load. Full-scale signal = mV/V × excitation.
  • 4-wire: E+, E−, S+, S−. Most common colours: red Ex+, black Ex−, green S+, white S−, though others are used.
  • 6-wire: adds Sen+ and Sen−, which let the instrument correct for cable resistance changing with temperature.
  • Bridge resistance: usually 350 Ω per arm; 120 Ω and 1,000 Ω also occur.
  • Multiple cells: wired in parallel (Ex+ to Ex+, S+ to S+), the signal is the average of all cells.
Protect the cell: Arc welding near a load cell can destroy its gauges. Wikipedia suggests disconnecting the cell and shorting all its pins to ground near the cell before welding nearby. Lightning can do the same damage.

Load Cell Output Calculator

Load Cell Output Calculator (mV/V)
IndustrialBasics calculator. Full-scale output = rated mV/V × excitation (Wikipedia, Load cell); output is taken as linear in load. With several identical cells wired in parallel through a junction box the signal is their average, so the reading is the total load over the combined capacity. An ideal estimate: real systems need zero and span calibration with known weights.

Types of Strain Gauge Load Cells

Types of strain gauge load cells and typical usesGrid of nine strain gauge load cell types: single-point, bending beam, shear beam, dual shear beam, S-type, compression, ring torsion, spoke or pancake, and load pin, each with its typical application.Load cell types and where they are usedSingle-pointplatform scales up to 1.2 × 1.2 mBending beampallet, platform, small hoppersShear beamlow-profile scales, 100 kg – 50 tDual shear beamtruck scales, tanks, hoppersS-typetension, static and dynamicCompressionweighbridges, large hoppersRing torsionhigh-accuracy silos, hoppersSpoke / pancakelow profile, 1 – 500 tLoad pinhoisting, dynamic forcesIndustrialBasicsindustrialbasics.com
Figure 2: Types and applications from Wikipedia (Load cell); IndustrialBasics graphic.

Beyond the shape, load cells differ in how load is applied. Multi-axis load cells measure force in more than one direction. Donut (annular) load cells are rings that fit over a bolt or shaft to measure clamping or press force. Silo load cells are normally compression or shear-beam cells under each leg, usually three for a vertical silo. Wikipedia notes that all load must pass through the sensing part of the cell. Friction or poor mounting can give offset, hysteresis or forces read along the wrong axis.

Other Load Cell Types

TypePrincipleStrengthsLimits
Strain gaugeResistance change in a bridgeAccurate, versatile, low costDamaged by overload and welding current; drift
HydraulicOil pressure under a piston and diaphragmNo electronics, hazardous areas, lightning-immuneCostlier; needs a pressure gauge or transmitter
PneumaticAir pressure balancing a diaphragmNo electrical partsNeeds an air supply
CapacitivePlate gap change, non-contact ceramic sensorOverload up to 1,000%, digital output over long cablesLess common than strain gauge
PiezoelectricCharge from deformationDynamic and impact forcesNot true static weighing
Vibrating wireWire frequency changes with tensionLow driftMainly geotechnical use
Principles and trade-offs from Wikipedia (Load cell).

Common Problems and Calibration

  • Overload: above rating, the body can deform permanently, causing zero offset, loss of linearity or a cell that cannot be calibrated.
  • Wiring: corrosion raising wire resistance, or moisture forming leakage paths, shifts the zero.
  • Temperature: thermal expansion and resistance changes look like strain; self-temperature-compensated gauges, dummy gauges and 6-wire sensing reduce it.
  • Nonlinearity at low load: cells used far below capacity need more calibration points.
  • Calibration: tare at zero, then apply known weights and set the span. SparkFun’s HX711 guide adjusts a calibration factor until the reading matches a known weight.
  • Reversed signal: readings that fall as load rises mean S+ and S− are swapped.
Integration: For PLC systems, a weighing indicator or signal conditioner converts the mV signal to 4–20 mA, 0–10 V or a fieldbus value. See types of industrial sensors for how load cells compare with other sensors, and 4–20 mA control loops for analog signal basics.

Frequently Asked Questions

What is a strain gauge load cell?

A force transducer made of a metal spring element with strain gauges bonded to it. Load deforms the element slightly, the gauges change resistance, and the resulting bridge voltage is proportional to the force (Wikipedia). It is the most common load cell in industry.

What is the difference between a load cell and a strain gauge?

A strain gauge is the sensing element: a fine foil or wire grid whose resistance changes with strain. A load cell is the complete transducer: a calibrated metal body, usually four gauges in a Wheatstone bridge, cable and sealing, rated in force or weight.

What other load cells exist besides strain gauge?

Hydraulic, pneumatic, capacitive, piezoelectric and vibrating-wire load cells (Wikipedia). Hydraulic cells have no electrical parts, which suits hazardous areas. Piezoelectric cells suit dynamic forces but not static weighing.

What is mV/V on a load cell?

The rated output: the signal in millivolts per volt of excitation at full rated load. A 2.96 mV/V cell gives 29.6 mV at full load with 10 V excitation (Wikipedia). Typical values are 1 to 3 mV/V.

What are the load cell wire colors?

Most commonly red Ex+, black Ex−, green S+ and white S− (Wikipedia), but other schemes exist. Always check the data sheet. If readings go down as load goes up, the signal wires are reversed (SparkFun).

Why do some load cells have 6 wires?

The extra two are sense wires that measure the excitation voltage at the cell, so the instrument can compensate for cable resistance changes with temperature (Wikipedia).

How many load cells does a tank or silo need?

Usually three for a vertical cylinder and four for a rectangular platform; long beams use two (Wikipedia). Replacing some cells with pivots saves cost but reduces accuracy.

What is a hydraulic load cell?

A piston and diaphragm filled with oil. Load raises the oil pressure, which a Bourdon gauge reads. With no electronics, it is immune to lightning transients and suits hazardous areas, but costs more (Wikipedia).

How do you calibrate a load cell?

Zero (tare) it with no load, then apply known reference weights and adjust the span or calibration factor until the reading matches. Use more points for wide ranges, since cells are less linear at the low end (Wikipedia).

What is an HX711?

A load cell amplifier IC sold on small breakout boards for microcontrollers such as Arduino. SparkFun’s guide connects E+, E−, A+ and A− to a four-wire load cell, then finds a calibration factor using a known weight.

Sources and Further Reading